Test distance :according to lab size
GON-1800T distribution photometer GON-1800T
Technical characteristics of Characteristics & Specifications
◆ principle fully meet the domestic and internationallatest relevant standards
Meet CIE, IESNA standard international testrequirements
The data format is according to international standard,the test results can be output in IESNAL international standard format, can be directly used as input data oflighting design software international.
Test data format meets international standard, can beoutput in IESNAL format and directly used in internationalilluminant
Design software
◆ adopts rotary lamp, B- β, A- α and C- γ etc. measuring met
By rotating luminaries to fulfill different measurementmethod such as B- β, A- α, C-
The use of international brands and domestic well-knownmotor reducer, transmission precision, the operation is extremely smooth
Use world-famous motors and Chinese reductors,precise transmission and running smoothly
◆ configuration grille lamps, cast light lamps, street lamps and other professional test fixture (other)
With professional test jigs for grille light, spot light andstreet light (Optional)
The innovative design of laser collimation adjustmentdevice of Innovation design for laser aiming calibrationequipment
◆ precision high sensitivity luminosity detector, standard grade V (λ) correctio
With accurate and high sensitivity photometer detector,standard fine V (λ) correctio
◆ RS-232 standard interface, Windows interface inEnglish
With RS-232 standard interface and English/ChineseWindows operation system
◆ with a handheld remote controller, easy to control two axis rotation
With handhold remote control, easy to control therotation of 2 shafts
◆ illuminance measurement range of Test range ofLuminance:0.001lx ~20000lx
◆ lamps around the vertical axis -180 ° ~180 ° or 0 °~360 ° rotation
Luminaries rotating around vertical axis:-180 ° ~180 ° or, 0 ° ~360 °
The lamp can rotate about a horizontal axis -180 ° ~180 ° or 0 ° ~360 ° rotation
Luminaries rotating around horizontal axis:-180 ° ~180 ° or, 0 ° ~360 °
The Angle accuracy: 0.1 ° angle accuracy
Maximum size of Max spec the of luminaries can bemeasuring fixture measured:
1) power of Power: 4kVA 2 Weight: 50kg 3) weight)dimensions of Outer dimension: 1500mm * 1400mm
◆ test distance: according to user laboratory size
Test distance: according to lab size